X ray facilities   -   GIXS

 

gixs 1 d384c
Description *
Manufacturer Enraf Nonius
Model FR591 (Rotating copper anode)
Goniometer Type : 3-circle
Company : Microcontrole
Rotation stages :
Theta, 2theta and phi
(all software controlled)
Translational/ Coarse control stages :
x , thetaY, thetaZ
(all manually controlled)
Detector
Type Scintillation detector with PMT
Company CRISMATEC
Crystal type  NaI (type 32 YB 2)
PMT Model XP 2072 B
Preamplifier Model As 17
Amplifier Model 590 A (EG & G Ortec)
Filament Tungsten
Fine Focus
Beam size on anode 0.1x1 mm2
Brilliance 12 kW/mm2
Focus spot type/dimension point focus/100 µm
High brilliance + high resolution
Beam size on anode 0.2x2 mm2
Brilliance  7.5 kW/mm2
Focus spot type/dimension point focus/200 µm
Broad focus
Beam size on anode 0.5x10 mm2
Brilliance
3 kW/mm2
High flux
Beam size on anode  1x10 mm2
Brilliance 1.5 kW/mm2
Focus spot type/dimension both vertical line focus as
  well as point focus
Beam size on anode 0.3x3 mm2
Brilliance 6 kW/mm2
Focus spot type/dimension point focus/300 µm
Target (Anode) Copper (optional Molybdenum) anode water cooled and vacuum sealed with
ferrofluid
speed : Min - 600 rpm ; Max - 6000 rpm
High Voltage Generator 2 separate modules 9 kW with oil cooled transformer
Prevacuum pump Type Leybold, TRIVAC 5BHV,
oil-sealed and air cooled,
capacity - 5 M3/Hr.
Final Vacuum pump Turbomolecular pump. Balzers,
TPH 240, 60,000 rpm,
ultimate pressure obtainable - 10(-7)

 